All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Abstract

Anodic oxidation of etodolac and its linear sweep, square wave and differential pulse voltammetric determination in pharmaceuticals

Author(s): B Yilmaz, S Kaban, BK Akcay
Department of Analytical Chemistry, Faculty of Pharmacy, Ataturk University, 25240 Erzurum, Turkey

Correspondence Address:
B Yilmaz Department of Analytical Chemistry, Faculty of Pharmacy, Ataturk University, 25240 Erzurum Turkey E-mail: yilmazb@atauni.edu.tr


In this study, simple, fast and reliable cyclic voltammetry, linear sweep voltammetry, square wave voltammetry and differential pulse voltammetry methods were developed and validated for determination of etodolac in pharmaceutical preparations. The proposed methods were based on electrochemical oxidation of etodolac at platinum electrode in acetonitrile solution containing 0.1 M lithium perchlorate. The well-defined oxidation peak was observed at 1.03 V. The calibration curves were linear for etodolac at the concentration range of 2.5-50 µg/ml for linear sweep, square wave and differential pulse voltammetry methods, respectively. Intra- and inter-day precision values for etodolac were less than 4.69, and accuracy (relative error) was better than 2.00%. The mean recovery of etodolac was 100.6% for pharmaceutical preparations. No interference was found from three tablet excipients at the selected assay conditions. Developed methods in this study are accurate, precise and can be easily applied to Etol, Tadolak and Etodin tablets as pharmaceutical preparation.

Full-Text | PDF

 
 
Google scholar citation report
Citations : 66710

Indian Journal of Pharmaceutical Sciences received 66710 citations as per google scholar report